Xi'an Kacise Optronics Co., Ltd.
TEL:86-29-88888109
Address::Tangyan South Road, Xi'an
Email: sales@kacise.com
Description
Electromagnetic flowmeter (abbreviated EMF) is the 1950s and 1960s with the development of electronic technology and the rapid development of a new type of flow measurement instruments. Electromagnetic flowmeter is the application of the principle of electromagnetic induction, the conductive fluid through the magnetic field induced electromotive force of an instrument to measure the flow of conductive fluid.
Specifications
Items | Description | |
Suitable Diameter | DN25~DN2600 (below DN25 non-standard) | |
Suitable Fluid | Liquid with conductive ratio more than 5 μs/cm | |
Measuring Range | 0.5~10m/s (extendable to 15m/s) | |
Upper Limit of Span | 0.5~10m/s. 1~5m/s recommended | |
Accuracy | 0.3% (Option), 0.5%, 1.0% of F.S.. | |
Repeatability | Below 0.25% | |
Display of Converter | Display : Flow rate ( 4-digits), Totalizer (9-digits), Velocity, Alarm status by LCD ( LED-option) | |
Scale unit: selectable of L/h, L/m, L/s, m3/h, m3/m, m3/s | ||
Flow direction : selectable by program | ||
3 Digits display below of decimal points. | ||
K-factor: 5 digits. | ||
Data logger for 2-3years. | ||
Programming language: English, Korean(on request) | ||
Mounting: Integral type; Remote Type | ||
Power | AC: 220V, 50-60Hz; DC: +24V(Option) | |
Output Signals | 4~20m ADC, load ≤ 750ohm, 0~3kHz, 5V with source and changeable width. High terminal and effective frequency output Voltage output:0-5 V DC | |
Conduit connection | M20 * 1.5 Thread | |
Communication Interface | RS-232; RS-485; HART | |
Operation Pressure | 1.0MPa, 1.6MPa, 4.0MPa, 16MPa (Special) | |
Fluid Temperature | -20°C~80°C, 80°C ~130°C, 130°C ~180°C Depending on Lining Material | |
Ambient Temperature | Sensor -40°C ~80°C. transducer -15°C ~50°C | |
Ambient Humidity | ≤ 85%RH (at 20C) | |
Cable Outlet Size | M20×1.5 | |
Power Supply | 220VAC±10%, 50-60Hz±1Hz, 24VDC±10%, Battery power (3-5years) | |
Power Consumption | ≤ 8W | |
Protection Ratings | Integral type: IP65. Remote type: sensor IP68, transducer IP65. | |
Materials | Electrode | 316L (stainless steel), Hastelloy C, Hastelloy B, Titanium, Tantalum, etc. |
Liner | PTFE, Soft rubber, Hard rubber, F46, FS, PUR. | |
Measuring tube | SUS 304 | |
Flange | Carbone Steal | |
Ground ring | 316L (stainless steel), HC, Ti, Ta, Cu. | |
Connection of Flange | National Standard GB9119-88 (DIN2051, BS4504), JIS, ANSI. Screw. Tri-clamp for sanitary application Insertion type. | |
EX-proof protection | Exde ib ii BT2 | |
Quality control | ISO9001-2008, CE |